Language:

EMG-EP Systems

Product Name : EMG-EP Systems
Model : EMG-EP Systems
 Features:
Professional hardware technology, performance is credible and steady
Multi processing unit parallel dispose technology, which makes data collection and graphics display more rapid
NueroScape EMG/EP has 4 channels, which satisfies different work models in the clinic EMG examination
Low noise, high CMRR, montage programmable electrode junction box
Independently programmable stimulators allows sophisticated examinations
Professional electromygram operation space, shortcut and steady
Intraoperative monitor capability
High speed and completely self-motion data collection and analysis, more nicety and more rapid with electrical stimulus, voice stimulus and light stimulus
USB 2.0 connection to PC or laptop

          
Main system part
  A/D conversion resolution: 16 Bit
  Sampling rate: 200kHz
  Analysis time: 5-5000ms
  Stimulating frequency: 0.1-50Hz
Amplifier part
  Four leads
  Sensitivity: 0.05μV-20mV/Grid
  Earth noise: EMG <= 4μV(Vpp)
  EP <=0.1μV(Vpp)(1000 times in average)
  Common-mode rejection ratio: >= 100dB
  50Hz Ban wave setting
  Upper limit Filter-frequency: 20kHz
  Lower limit Filter-frequency: 0.01Hz
  Plus amplifies 25 times-400000: times
Stimulator part
  Current stimulator:
  constant current 0.2-100mA
  pulse width 50-1000μS short circuit and over
  loading protection