Multi-Parameter Monitor PM900A

Product Name : Multi-Parameter Monitor PM900A
Model : PM900A

Ÿ  15’ TFT display

Ÿ  Unique human voice alarm

Ÿ  720 hours trend data and 2hours of all waveforms

Ÿ  Wireless network

Ÿ  14 multi-languages include:English, French, German, Spanish, Portuguese, Russian etc

Ÿ  Function: ECG, RESP, TEMP, NIBP,SPO2. PR

Optional:

Ÿ  Dual IBP CO2 CSM/IOC

Ÿ  Multi-gas and fetal monitorRecorder

Ÿ  Wire network

Ÿ  Touch Screen


 

ECG

Full lead: I, II, III, AVL, AVR, AVF and V
Gain: 5mm/mV, 10mm/mV, 15mm/mV, 20mm/mV and 25mm/mV
Scanning speed: Changeable with three grades of 12.5mm/s, 25mm/s and 50mm/s
Band width: 0.05Hz-100Hz(3dB)
Heart rate: 
Scope: 15 times/minute to 300 times/minutes
Accuracy:  ±1 time/minute
Variation of heart beat LORENZE drawing: Analysis on heart beat data for the current 5 minutes with histogram during RR period. 
Arrhythmia: Twenty kinds of common abnormal ECG can be analyzed, with ECG data of 1000 times of arrhythmia kept and with 10 seconds kept for each time.   
Accuracy: Error rate < 20%         
Undetected rate < 20% 
ST section 
Scope:     ±0.8mV
Precision :  < ±0.05mV

SpO2

Detection scope:   0%-100%
Accuracy         ±1%     90%-100%               
                 ±2%      70%-89% 

NIBP 

Theory Electric oscillator 
Cuff gassing time   less than 15 seconds
First-time gassing pressure: dynamic state 24kPa(180 mmHg)
Manual selection: 9.3kPa(70 mmHg),13.3kPa(100 mmHg),16.0kPa(120 mmHg)
18.6kPa(140 mmHg),20.0kPa(150 mmHg),21.3kPa(160 mmHg),
Automatic detection period: 1 minute to 240 minutes 
Stop size:1 min(1 min -10min), 5min(10 min -30min), 10min(30 m
in -90min) and 30min(90 min -240min)
Detection scope: 
Systolic blood pressure:   4kPa -34kPa(30mmHg -255mmHg)
Diastolic pressure             2kPa -29.3kPa(15mmHg -220mmHg)
Mean pressure               2.7kPa~31.3kPa       (20mmHg~235mmHg)
Detection accuracy 
Static pressure: ±3mmHg
Pulse rate:  ±2%

Resp

Chest impedance method or concentration of carbon dioxide at end of respiration 
Scope of chest impedance method     0 time/minute – 100 minutes/minute 
Precision of chest impedance method  time/minute 
Detection scope of carbon oxygen at end of respiration 3mmHg-75mmHg; 0.4%vol-9.9%vol
Detection scope of respiration rate: 2 times/minute ->60 times/minute(including baby)
Detection precision of respiration rate: ±1 time/minute 

Temp

Single/double path 
Type:    Vascular/surface temperature 
Scope:    20℃~50℃
Accuracy:  ±0.2℃

Invasive blood pressure (optional)

Single/double path 
Type: direct and continuous detection 
Scope: -6.67kPa(-50mmHg)—46.67kPa(350mmHg)
Precision: below 13.33kPa(100mmHg) ±0.27kPa(2mmHg)  13.33k
Pa(100mmHg) above ±2%
Frequency response:DC-12Hz
Input resistance: not less than 500kΩ
Sensitivity of sensor: 5μV/V/mmHg
Motivation voltage:5V (DC)
Accuracy of sensor: ±2%Isolation voltage: 4 kV